The GIT measurements of CH2O, HONO, and NO will
be an important part of the ANTCI effort to gain a quantitative understand
of HOx chemistry
as well as to understand the reactive nitrogen budget. Previous
polar studies have shown that CH2O,
HONO, and NO must be measured if models are to be comprehensively tested
in polar regions. The Georgia Tech
Laser-induced fluorescence group will simultaneously measure formaldehyde
and nitrous acid during the 2003 field study at the South Pole, and
formaldehyde and nitric oxide (NO) during the 2005 airborne campaign.
Formaldehyde will be measured by direct excitation at 353 nm with detection
occurring at ~ 430 nm. The excitation wavelength has been chosen such
that the photodissociation yield is minimized and interference avoided.
It is generated by frequency doubling the output of a 706 nm optical
parametric oscillator, pumped by the second harmonic of a YAG laser.
Ambient air is drawn into the sample cell through a small orifice by
means of a Roots blower and vacuum pump. The current laboratory version
of this system has a detection limit of 120 pptv; however, the final
optimized field system is expected to have a 2-s LOD of less than 5
pptv. Calibrations will be by standard addition to the flow using a
formaldehyde permeation tube maintained at 60 oC as a source.
Calibrations and other system diagnostic checks will be performed on
an hourly basis
on the ground, and as the flight path requires in the air.
The GIT NO measurement is based on two-photon laser-induced fluorescence
(TP-LIF) detection and has been used in numerous airborne field studies
which have included intercomparisons with other techniques. This system
uses two excitation wavelengths, one at 226 nm exciting the NO molecule
into
the A state, and a second wavelength at 1097 nm, which pumps NO from the
A to the D state. Fluorescence in this system is therefore blue shifted
with major emission occurring at 187 nm. Thus, the 2-photon technique allows
for very high background rejection, making the measurement essentially
background-free. A typical NO LOD using TP-LIF is 1 pptv or better. In
previous field campaigns, NO and NO2 have been measured using 2 YAG lasers
and 2 large master-oscillator/power-oscillators (MOPOs). Improved efficiencies
in laser light generation and plus improvements in detection will allow
us to reduce this payload to a single YAG laser, plus several small OPOs.
Detection will take place in a multipass cell behind the CH2O detection
cell. Calibration will be done using standard addition techniques based
on a NIST-traceable NO standard, and will be done on a routine basis during
the flight.
HONO will be measured by photo fragmentation using the third harmonic of
a YAG laser (e.g., 355 nm) followed by LIF detection of the OH moiety in
a cell located downstream from the CH2O detection cell.
The OH fragment will be excited at 308 nm, and the time-gated resonance
fluorescence detected on microchannel plates. This technique has also been
proven on many field campaigns dating back to 1996. The OH limit of detection
for this technique has been shown to be better than 0.01 pptv. With 2W
of photolysis power, the
photolytic efficiency is calculated to be better than 10%. Thus, assuming
a typical 0.2 to 0.3 pptv ambient OH signal, the HONO limit of detection
is conservatively estimated at ~1 pptv. Calibration will be done by standard
addition of HONO. As in the case of CH2O, calibrations
and other system diagnostics will be performed on an hourly basis.
|